PFA for Atrial Arrhythmias

Definition

Pulsed field ablation for atrial arrhythmias is the delivery of high-voltage, short-duration electrical pulses to atrial myocardium to achieve irreversible electroporation — in practice, durable pulmonary vein isolation for atrial fibrillation, with extensions to the posterior wall, superior vena cava, mitral isthmus, left atrial appendage and cavotricuspid isthmus. It is the first ablation energy that does not depend on heating or freezing tissue, and the first in two decades to change the safety ceiling of the procedure: more than 500,000 pentaspline procedures have been performed worldwide without a single reported atrio-oesophageal fistula, against a thermal incidence of 0.025% and 66% mortality. (sources/pfa-af-ehra-2026, rating: very high) The ventricular application is a separate problem with a different evidence base and its own constraints — see concepts/Ventricular-PFA.

Key Concepts

Mechanism and what selectivity means

Ablation method Cell damage Microvasculature Extracellular matrix Healing
Radiofrequency Indiscriminate protein denaturation Thrombogenic, indiscriminate thermal damage Coagulation Delayed, from periphery
Cryoablation Crystal formation, membrane damage, osmotic imbalance Clogged and damaged Preserved Delayed, from periphery
Pulsed field Membrane lipid oxidation and protein damage Preserved, with transient hypoperfusion and increased permeability Preserved Facilitated by preserved microcirculation

(Table: sources/pfa-af-ehra-2026)

Device classification and platforms

The consensus proposal classifies catheters by the largest dimension in contact with tissue, replacing "single-shot", "wide-area focal" and "lattice-tip":

Class Definition Former terms
Regional Contact dimension >12 mm Circumferential PVI catheter; single-shot
Large-tip focal >4 to ≤12 mm Wide-area focal; lattice-tip (Affera); Omnipulse
Focal ≤4 mm Conventional focal; large/intermediate footprint

Functionally the systems split three ways: circumferential or large-area devices for rapid PVI (simplest, best safety data); large-footprint catheters integrated with 3D mapping (adaptable to linear and ventricular lesions, but mapping-dependent); and contact-force RF catheters repurposed to deliver PFA (versatile and familiar, but point-by-point). (sources/pfa-af-ehra-2026)

System (manufacturer) Architecture Size, diameter Mode Amplitude Delivery
Farapulse (Boston Scientific) Pentaspline, over the wire 12 F, 31/35 mm Bipolar biphasic 1.8–2.0 kV/µs 2.5 s; ≥8 per PV advised
PulseSelect (Medtronic) Circular array, over the wire 9 F, 25 mm Bipolar biphasic 1.5 kV/µs 4–5 s; ≥8 per PV advised
Varipulse (Biosense Webster) Variable-loop circular 8.5 F, 25–35 mm Bipolar biphasic 1.8 kV/µs ~22 s per ablation (3 applications); 4 per PV
Volt (Abbott) Basket, impedance contact sensing 13 F, 28 mm Bipolar biphasic 1.5–1.8 kV/µs ~20 s, 10 R-wave-gated bursts
Globe PF (Kardium) Spherical array 16 F, 30 mm Bipolar biphasic 1.7 kV/µs 2 s per train, 2–6 trains
Sphere 360 (Medtronic) Spherical array 8 F, 34 mm Monopolar biphasic 1.3–2.0 kV/µs 5–6 s; 4 per PV
Sphere-9 (Medtronic) Lattice tip, 3D integrated, dual energy 8 F, 9 mm Monopolar biphasic 2 kV/µs 4 s
Omnypulse (Biosense Webster) Large-tip focal, 3D integrated 7.5 F, 12 mm Bipolar biphasic 1.2–1.5 kV/µs Up to 12 applications
Faraflex (Boston Scientific) Large-tip focal, 3D integrated 8 F, 9 mm Monopolar/bipolar 2 kV/µs 2.5 s
QuickShot (CardioFocus) Large-tip focal 8.5 F, 10 mm Monopolar biphasic 3.5 kV/µs 3 s (QS) / 6.8 s (QS+)
Farapoint (Boston Scientific) RF-style focal 8 F Bipolar biphasic 1.4–2.0 kV/µs 2.5 s
Centauri (CardioFocus) RF-style focal, contact force sensing 8 F Monopolar biphasic 19–25 A/µs 3–7 s by current
STSF Dual Energy (Biosense Webster) RF-style focal, CF sensing, dual energy 8 F Monopolar biphasic 1.0–1.5 kV/µs Up to 24 applications
TactiFlex Dual Energy (Abbott) RF-style focal, CF sensing, dual energy 8 F Monopolar biphasic 1.9–2.2 kV/µs ~5 s (2.2 kV) / ~10 s (1.9 kV)

Adagio, Lotos, nsPFA, Optishot and Arga systems are in development; no complete list is possible because devices change continuously, and waveform parameters are rarely disclosed, which the writing committee names as a barrier to comparison. (sources/pfa-af-ehra-2026)

Energy delivery and dosing

Dual-energy systems

A single catheter and generator switching between RF and high-voltage PF trains under unified power management, with impedance, temperature and voltage feedback. The rationale is complementarity — PFA for thin-walled structures, RF for thicker myocardium and where coronary spasm is a concern, and combined PF+RF on one trajectory ("lesion stacking") for greater depth. Named systems are the Affera Sphere-9 and the ThermoCool SmartTouch SF Dual Energy with TRUPULSE. Comparative outcome data against single-energy PFA are awaited. (sources/pfa-af-ehra-2026)

Anaesthesia

Imaging, mapping and ICE

Device interactions

Step-by-step

  1. Access: ultrasound-guided femoral venous puncture; uninterrupted oral anticoagulation where possible; ACT ≥300–350 s (≥350 s before left atrial manipulation for some systems).
  2. Guidewire: a standard J-tip suffices for over-the-wire devices; extra-stiff wires add support and perforation risk; straight-tip wires have been associated with asymptomatic bronchial bleeding.
  3. Transseptal puncture: posterior is standard, anterior can ease right inferior PV access when the LA is not enlarged; a single over-the-needle puncture with the PFA sheath ("zero-exchange") has been described in 100 patients.
  4. Air embolism prevention: meticulous de-airing, continuous saline flush, and zero-exchange designs that also shorten fluoroscopy and LA dwell time. Negative LA pressure under sedation may favour air ingress.
  5. Vagal response: 30–70% incidence, mostly at the left superior PV, usually self-limited; atropine 1.0 mg, rarely temporary pacing. Prophylactic parasympatholytics are contraindicated in glaucoma and benign prostatic hyperplasia. RSPV-first ablation reduced vagal reactions from 78% to 13%.
  6. Contact: coaxial antral alignment, never deep in the vein; assess by tactile feedback, fluoroscopic indentation, pacing capture, ICE and impedance-based mapping markers (good contact gave 100% transmurality in vivo). PFA is less forgiving of intermittent contact than RF, because ultra-short delivery has no thermal latency and proximity must exist at the instant of delivery.
  7. Projections: LAO for left veins, RAO for right veins; biplane may help.
  8. Interruption troubleshooting: automatic cut-off occurs with unsafe geometry, abnormal impedance or metallic interaction — touching splines, loop diameter below the validated range, a wedged lattice tip (temperature overshoot) or balloon malposition. Arcing is rare and requires electrode–electrode or electrode–metal contact.
  9. Endpoint: near-field/far-field discrimination is harder after PFA and varies with electrode size and spacing; durable entrance and exit block remain the universal endpoint, adenosine has low yield, and a standard waiting time adds little.
  10. Post-procedural care: uninterrupted anticoagulation ≥2 months, then by CHA₂DS₂-VA; most centres have abandoned routine proton-pump inhibition; same-day discharge is supported in selected patients by registries, an admIRE sub-analysis and an EHRA meta-analysis, and is inappropriate after complications.

(All: sources/pfa-af-ehra-2026)

Training and the learning curve

Clinical Evidence

The five randomised trials

Trial Population PFA vs comparator N Freedom from AF/AT Safety endpoint Procedure time
ADVENT Paroxysmal Pentaspline vs thermal 607 73.3% vs 71.3% 2.1% vs 1.5% 106 vs 123 min
SINGLE SHOT CHAMPION Paroxysmal Pentaspline vs cryoballoon 210 62.9% vs 49.3% 1.0% vs 1.9% 55 vs 73 min
BEAT PAROX-AF Paroxysmal Pentaspline vs RF 289 77.2% vs 77.6% 4.8% vs 7.6% 56 vs 95 min
INSIGHT PFA Paroxysmal Nanosecond biphasic vs RF 287 66.7% vs 67.4% 18.3% vs 17.9% 99 vs 113 min
SPHERE Per-AF Persistent Lattice-tip dual-energy vs RF 420 73.8% vs 65.8% 1.4% vs 1.0% 101 vs 126 min

Trial detail worth retaining

Registries and non-randomised comparisons

Registry N Major adverse events Tamponade Death Stroke Persistent PNI Oesophageal fistula
MANIFEST-PF 1,568 1.9% 1.1% 0.06% 0.4% 0 0
MANIFEST-17K 17,642 0.9% 0.36% 0.03% 0.12% 0 0
EU-PORIA 1,233 1.7% 1.1% 0 0.41% 0 0

Persistent AF — the weakest evidence

Redo procedures

Blanking period

Applications Beyond Pulmonary Vein Isolation

Target Status Key limitation
Posterior wall Feasible with pentaspline, lattice-tip and spherical-array systems; transmurality verified by epicardial mapping; no oesophageal injury reported No 12-month benefit with adjunctive ablation in MANIFEST-PF; PIFPAF-PFA testing prospectively
Superior vena cava Achievable with pentaspline and circular arrays Empirical vs targeted isolation untested; safety may differ by platform
Mitral isthmus High acute block rates, especially with the dual-energy lattice tip (68% durability at remapping; 1.4% adverse events in SPHERE Per-AF) 4–10% clinical and up to 40% subclinical coronary spasm; durable block only 5.5% at appendage-occlusion remapping with pentaspline
Left atrial appendage isolation Feasible (~16 applications) with low complication rate 95% reconnection at 3 months
Scar homogenisation 100% acute success, 64% freedom at 6 months in one focal-PFA series No randomised data
CTI-dependent flutter 99.5% acute success with pentaspline off-label; 100% with lattice tip (mostly RF) Right coronary artery spasm, occasionally with malignant ventricular arrhythmias; use IV nitroglycerine
PSVT / AVNRT High acute and 6-month success in Chinese focal-PF cohorts Transient AV block in up to 17.5% with one bipolar focal catheter

(All: sources/pfa-af-ehra-2026, rating: very high)

Safety

Event Reported range with PFA Comparator
Oesophageal fistula 0 in >500,000 procedures Thermal 0.025% incidence, 66% mortality
Persistent phrenic nerve injury 0% (transient 0.006%, MANIFEST-17K) Thermal 0–1%
Severe PV stenosis Not observed with PFA-only devices Frequent mild–moderate narrowing with thermal
Tamponade 0–1.1% Similar
Stroke / TIA 0–1% / 0–0.3% Similar (0.1–1% combined)
Silent cerebral events 14% RF 18%, cryoballoon 21%
Vascular access complications 0–2.2% (two-thirds lower with ultrasound) Similar
Death 0–0.3% Thermal 0.02–0.46%
Pericarditis 0–0.3% (up to 4.4% when actively sought) Thermal 0–10.2%
Coronary vasospasm 0.1–0.14% in AF registries; far higher for CTI, mitral isthmus and epicardial ventricular delivery Not a thermal phenomenon
Haemolysis / AKI Application-number and voltage dependent Not a thermal phenomenon

(Table: sources/pfa-af-ehra-2026, rating: very high)

Platform-specific and delayed concerns