Self-Expanding or Balloon-Expandable TAVR in Patients with a Small Aortic Annulus (SMART)

Authors, Journal, Affiliations, Type, DOI

Overview

SMART randomised patients with symptomatic severe aortic stenosis and an aortic annulus area of 430 mm² or less to a self-expanding supraannular valve (Evolut PRO/PRO+/FX) or a balloon-expandable valve (SAPIEN 3/3 Ultra) — a population that is about 87% women and historically under-represented in device-versus-device comparisons. At 12 months the self-expanding valve was non-inferior for the clinical composite of death, disabling stroke or heart-failure rehospitalisation (9.4% vs 10.6%), and superior for a composite of bioprosthetic-valve dysfunction (9.4% vs 41.6%). The difference in that second end point is almost entirely haemodynamic — a mean gradient of 7.7 vs 15.7 mmHg — so the trial establishes better valve performance in small annuli at one year, not better clinical outcomes or better long-term durability.

Keywords

Aortic stenosis; small aortic annulus; transcatheter aortic-valve replacement; TAVR; self-expanding valve; balloon-expandable valve; prosthesis–patient mismatch; bioprosthetic-valve dysfunction; valve haemodynamics

Key Takeaways

Background

Methods

Results

Participants

Coprimary End Points

End point Self-expanding Balloon-expandable Difference (CI)
Death, disabling stroke or HF rehospitalisation at 12 mo 9.4% 10.6% −1.2 pp (90% CI −4.9 to 2.5); P<0.001 for non-inferiority; HR 0.90 (0.56–1.43)
— Death 5.1% 5.9% −0.7 pp (−3.5 to 2.1)
— Disabling stroke 3.1% 2.6% 0.6 pp (−1.5 to 2.7)
— HF rehospitalisation 3.8% 3.5% 0.4 pp (−2.0 to 2.7)
Bioprosthetic-valve dysfunction at 12 mo 9.4% 41.6% −32.2 pp (95% CI −38.7 to −25.6); P<0.001 for superiority
— Haemodynamic structural valve dysfunction 3.2% 32.2% −29.1 pp (−34.6 to −23.5)
— Nonstructural valve dysfunction 5.9% 18.2% −12.3 pp (−17.6 to −7.0)
— Clinical valve thrombosis 0.3% 0.3% 0.0 pp (−0.8 to 0.8)
— Endocarditis 0.6% 2.3% −1.7 pp (−3.5 to 0.1)
— Aortic-valve reintervention 0.9% 0.6% 0.3 pp (−1.0 to 1.6)

Secondary and Additional End Points

End point Self-expanding Balloon-expandable Difference (95% CI)
Mean gradient at 12 mo 7.7 mmHg 15.7 mmHg −8.0 (−8.9 to −7.1)
Effective orifice area at 12 mo 1.99 cm² 1.50 cm² 0.49 (0.42 to 0.56)
Haemodynamic structural valve dysfunction 3.5% 32.8% −29.3 pp (−34.7 to −23.8)
Bioprosthetic-valve dysfunction in women 10.2% 43.3% −33.0 pp (−40.2 to −25.9)
Moderate or severe prosthesis–patient mismatch at 30 d 11.2% 35.3% −24.1 pp (−30.5 to −17.7)

Safety

Subgroups

Discussion

Limitations of the document