Non-Invasive Assessment of Ventricular Electrical Heterogeneity to Optimize Left Bundle Branch Area Pacing

Authors, Journal, Affiliations, Type, DOI

Overview

This study used a 40-electrode body-surface ECG belt to characterise ventricular electrical heterogeneity (VEH) during different forms of left bundle branch area pacing (LBBAP) in 20 patients, most with a CRT indication. It measured global heterogeneity as the standard deviation of activation times (SDAT) across all belt electrodes, plus average LV activation time (LVAT) and separate LV and RV dispersion. Optimised LBBAP reduced every metric substantially, with a 47% fall in SDAT but only a 16% fall in QRS duration, and produced ~80% clinical and echocardiographic response at ~8 months. The study's most distinctive findings are that the electrical benefit was independent of capture mechanism — selective, nonselective, anodal and LV septal capture could not be separated — and that anodal capture worsened LVAT and LV dispersion despite shortening QRS.

Keywords

Cardiac resynchronization therapy; ECG belt; Left bundle branch pacing; LV septal pacing; Anodal capture

Key Takeaways

Background and rationale

Definitions used

Methods

Baseline characteristics

Procedural characteristics

ECG belt results

Metric Optimised LBBAP Intrinsic Change
QRS duration 133 ± 20 ms 157 ± 24 ms −16%
SDAT 20.5 ± 7.5 ms 38.6 ± 9 ms −47%
LVAT 44.4 ± 14.3 ms 61.4 ± 21 ms −28%
LV dispersion 11.6 ± 11.6 ms 29.5 ± 15 ms −61%
RV dispersion 21.1 ± 7.8 ms 42.5 ± 9.3 ms −50%

Echocardiographic and clinical outcomes

Correlation between SDAT reduction and response

Discussion points

Limitations of the document