Safety, Efficacy, and Functional Outcomes Using a Novel Small-Diameter Defibrillation Lead for LBBAP: Final Results of LEADR LBBAP

Authors, Journal, Affiliations, Type, DOI

Overview

LEADR LBBAP evaluated the 4.7-F, lumenless, catheter-delivered, integrated bipolar, single-coil OmniaSecure defibrillation lead (Medtronic) when implanted in the left bundle branch area. Of 316 implantation attempts, 302 (95.6%) succeeded; 157 patients received LBBAP-ICD therapy and 143 received LBBAP-optimized CRT (LOT-CRT). Over a mean follow-up of 9.9 months, electrical performance was stable (pacing threshold, impedance and R-wave amplitude), freedom from study lead–related major complications was 97.7% at 6 months and 96.8% at 12 months, and appropriate therapy was delivered in 7.9% of patients at 1 year with a 2.1% inappropriate-shock rate. In the LOT-CRT subgroup, QRS narrowed from 170.6 to 157.5 ms, LVEF rose from 29.7% to 38.1%, LVESVi fell from 53.8 to 40.5 mL/m², and KCCQ improved by 14.2 points at 6 months. The trial supports a single-lead strategy that provides defibrillation and physiological pacing without a separate RV pace/sense lead.

Keywords

Conduction system pacing; Left bundle branch area pacing; Defibrillation lead; Cardiac resynchronization

Key Takeaways

Study design and population

Implantation and electrical performance

Safety

Arrhythmia therapy

LOT-CRT functional outcomes (n = 143)

Interpretation