Initial Experience of Left Bundle Branch Area Pacing Using Stylet-Driven Pacing Leads: A Multicenter Study

Authors, Journal, Affiliations, Type, DOI

Overview

This registry asks whether left bundle branch area pacing (LBBAP) can be performed with ordinary stylet-driven pacing leads (SDL) rather than the specialised lumenless lead (Medtronic SelectSecure 3830) that nearly all published LBBAP experience had used. Between eight Belgian hospitals, 353 consecutive patients were implanted with the Solia S60 (Biotronik, 5.6 Fr, extendable helix) through a dedicated Selectra3D sheath, with implant success defined as either left bundle branch pacing or left ventricular septal pacing. Success was 94.6%, unipolar thresholds were low (0.6 ± 0.4 V at 0.4 ms) and stable to 12 months, and — the study's most consequential finding — success was independent of centre volume (93–100% across centres, p = .75), including centres with as few as 5 implants. The safety signal cuts the other way from the lumenless literature: septal perforation (2%) and septal coronary artery fistula (1.4%) occurred at rates above those reported for lumenless leads, though all perforations were asymptomatic and no ventricular septal defect or septal haematoma persisted. Its central limitation is definitional — only 73% of "successful" implants achieved actual left bundle capture, so the remaining 27% were pure myocardial (LVSP) pacing, a looser criterion than the reports it is implicitly measured against.

Keywords

left bundle branch area pacing; left ventricular septal pacing; physiologic pacing; stylet-driven pacing leads

Key Takeaways

1 | Introduction

2 | Methods

2.1 | Study population

2.2 | LBBAP via stylet-driven pacing leads — the technique

The preparation steps are the practical core of the paper, because an extendable helix must be prevented from retracting while the outer lead body is rotated:

2.3 | Defining successful LBBAP

2.4 | Data collection and follow-up

2.5 | Safety endpoints and complications

2.6 | Statistical analysis

3 | Results

3.1 | Patient characteristics

3.2 | Procedural success and characteristics of LBBAP using SDL

Parameter Value
Implant success 334 (95%) — see the reporting-inconsistency note in Limitations
Screw attempts, n 2.2 ± 1.9
Implant depth, mm 13 ± 2
Procedural time, min 60 [60–80] (median [IQR])
Total fluoroscopy time, min 7 [4–11] (median [IQR])

3.3 | Electrophysiological and pacing characterization of LBBAP using SDL

Parameter Value
Paced QRS duration, ms 126 ± 21
Stimulus to LVAT, ms 74 ± 17
LBB potential, n (%) 48 (14%)
NS-LBBP 192 (57%)
S-LBBP 52 (16%)
LVSP 90 (27%)
Unipolar pacing threshold at 0.4 ms, V 0.6 ± 0.4
Bipolar pacing threshold at 0.4 ms, V 0.9 ± 1.7
Unipolar R-wave amplitude, mV 10 ± 5
Bipolar R-wave amplitude, mV 11 ± 5
Unipolar impedance, ohms 441 ± 106
Bipolar impedance, ohms 607 ± 94
Baseline Pre-LBBAP QRS Post-LBBAP QRS
LBBB 152 ± 20 ms 130 ± 20 ms
RBBB 143 ± 19 ms 130 ± 23 ms
IVCD 128 ± 20 ms 125 ± 21 ms
Narrow QRS (<120 ms) 95 ± 16 ms 125 ± 18 ms (increased)

3.5 | Echocardiographic outcome

Complication Patients (n)
Complications at implant
Pneumothorax 4
Hemothorax 1
Tamponade related to atrial lead 2
Tamponade related to LBBAP lead 0
Takotsubo cardiomyopathy 1
Permanent RBB injury 1
New persistent iatrogenic AV block 0
Septal perforation 8 (2%)
Septal coronary artery fistula 5 (1.4%)
Helix fracture or elongation resulting in lead disuse 3
Complications during follow-up
Revision for pocket hematoma 1
Atrial lead revision 3
LBBAP lead revision at day 1 2
— reason: lead dislodgement to right ventricle 2
— reason: septal perforation 0
LBBAP lead revision beyond day 1 3
— late septal perforation 1
— lead dislodgement to right ventricle 1
— lead fracture (impedance >2000 Ω) 1
Echocardiographic follow-up
SCAF persisting >3 months 0
Septal hematoma 0
Ventricular septal defect 0

3.6 | Implant success in low and high-volume centres — the study's key finding

3.x | Pacing characteristics at follow-up (Table 3)

Pacing characteristic Implant 1 Month 6 Months 12 Months p value
Number of patients, n 353 349 259 67
Pacing threshold at 0.4 ms, V 0.6 ± 0.4 0.6 ± 0.2 0.7 ± 0.2 0.7 ± 0.3 .291
R-wave sensing, mV 11 ± 5 12 ± 5 12 ± 5 13 ± 6 .061
Unipolar impedance, ohms 441 ± 106 399 ± 90 384 ± 74 397 ± 75 .115

4 | Discussion

4.1 | Main findings

4.2 | Current experience with LBBAP using LLL and SDL

4.3 | Safety profile of SDL used for LBBAP

4.4 | Advantages and disadvantages of SDL used for LBBAP

4.5 | Limitations (authors' own — see also the section below)

5 | Conclusion

Limitations of the document

Authors' own stated limitations

Additional limitations evident in the study

Key Concepts Mentioned

Key Entities Mentioned

Wiki Pages Updated