Left Ventricular Lead Placement Guided by Reduction in QRS Area

Authors, Journal, Affiliations, Type, DOI

Overview

This study asks whether the reduction in QRS area during biventricular pacing can guide LV lead placement, by comparing it with the acute haemodynamic response (%ΔLVdP/dtmax) across different pacing sites within the same patient. In 52 patients and 188 pacing measurements pooled from two prior studies, lateral and anterolateral LV pacing produced both the largest QRS-area reduction and the largest haemodynamic improvement, while QRS duration did not change across sites. Within patients, ΔQRS area correlated strongly with %ΔLVdP/dtmax (median R 0.76), and the largest QRS-area reduction coincided with the largest haemodynamic gain in the great majority. The authors propose QRS area as an objective, non-invasive intraprocedural tool for choosing the LV pacing site and the active electrode on a quadripolar lead.

Keywords

Cardiac resynchronization therapy; LV lead placement; QRS area; invasive hemodynamic measurements; heart failure

Key Takeaways

Background and rationale

Methods

Baseline characteristics

QRS-area reduction and haemodynamics by anatomical site

In 26 patients with measurements at different anatomical LV lead positions:

LV pacing site ΔQRS area %ΔLVdP/dtmax
Lateral −48 µVs 24%
Anterolateral −43 µVs —
Posterolateral −17 µVs —
Anterior −13 µVs 12%
Posterior +9 µVs 11%

Proximal vs distal electrode pair

Relationship between ΔQRS area and %ΔLVdP/dtmax

Interpretation

Limitations of the document

Key Concepts Mentioned

Key Entities Mentioned

Wiki Pages Updated