Safety and Efficacy of Catheter Ablation for Atrial Fibrillation in Patients with Percutaneous Atrial Septal Closure Device

Authors, Journal, Affiliations, Type, DOI

Overview

Patients with a percutaneous atrial septal defect (ASD) occluder who develop AF present a specific technical problem: transseptal puncture for left atrial access must be made through, or around, an implanted device, and the perceived risk of device dislodgement has historically pushed electrophysiologists toward long-term antiarrhythmic drugs instead. This meta-analysis pooled the only three published series — 64 patients in total, all from high-volume centres — and found that puncture succeeded in 100%, freedom from AF was 77.7% (95% CI 65.7–86.3) over 6–22 months, and no device was dislodged and no residual interatrial shunt developed. Puncture through the device took significantly longer than through native septum (237.3 versus 180 minutes total procedure time, P=0.004), and recurrence did not differ between approaches. The price of the technique as reported is a 4.7% rate of inadvertent epicardial (aortic root) puncture, which the authors flag as high even for expert operators. The evidence base is three uncontrolled single-centre series, which is the main reason this page rates the paper medium rather than high.

Keywords

atrial fibrillation ablation; atrial septal device closure; transseptal puncture

Key Takeaways

The clinical problem

Why AF matters in this population

Search and included studies

Patient characteristics

Procedural results

Comparison Native septum Through device P
Recurrence of AF 23.07% 16.66% 0.79 (RR 1.18, 0.35–4.00)
Total fluoroscopy time 43.50 min 70.67 min 0.44
Total procedure time 180 min 237.3 min 0.004

Complications

Why there is no residual shunt

Timing and extension to other left-sided procedures

Limitations of the document

Authors' own stated limitations

Additional limitations evident in the design