Electrophysiological Characteristics and Possible Mechanism of Bipolar Pacing in Left Bundle Branch Pacing

Authors, Journal, Affiliations, Type, DOI

Overview

Left bundle branch pacing is programmed bipolar as often as unipolar, yet its capture physiology had been described almost entirely in unipolar terms. This study performs a decremental bipolar threshold test in 65 patients with strictly confirmed LBB capture and finds that bipolar pacing passes through two transitions, not one. At high output, the cathode captures the LBB while the anode captures the right ventricular septum — a state the authors call NS-bipolar-LB — producing the shortest paced QRS (118.2 ms) and shortest V1 RWPT (95.6 ms) of any capture type. As output falls, anodal capture is lost, giving NS-cathodal-LB (QRS 133.8 ms), and then either selective cathodal LBB capture (S-cathodal-LB) or LVS-only capture depending on the patient. V6 RWPT stayed short and constant across all three LBB-capturing states (64.8–65.7 ms) and lengthened only with LVS-cathodal capture (82.8 ms). The clinical speculation is that high-output bipolar pacing's simultaneous right and left septal capture may be a more synchronous mode than conventional unipolar LBBP — and that bipolar should be preferred for LBBP because its low-output behaviour is identical to unipolar while its high-output QRS is narrower.

Keywords

Physiological pacing; Left bundle branch pacing; Bipolar pacing; Intracardiac electrogram; Continuous pacing technique

Key Takeaways

Population and capture verification

The two bipolar transitions

The four bipolar capture types

Type Mechanism proposed V6 RWPT V1 RWPT Paced QRS Threshold
NS-bipolar-LB Cathode captures LBB and anode captures RV septum 64.8 ± 7.7 ms 95.6 ± 10.3 ms 118.2 ± 14.2 ms 2.5 ± 1.2 V
NS-cathodal-LB Anodal capture lost; cathode captures LBB + local LVS myocardium 65.7 ± 7.8 107.5 ± 12.7 133.8 ± 15.8 0.8 ± 0.4 V
S-cathodal-LB Selective cathodal LBB capture 65.7 ± 7.3 120.5 ± 15.2 145.3 ± 16.8 0.5 ± 0.2 V
LVS-cathodal LBB capture lost; cathode captures LVS only 82.8 ± 8.6 112.1 ± 10.1 140.9 ± 11.5 0.6 ± 0.1 V

The proposed mechanism

Clinical implications the authors propose

Limitations of the document

Authors' own stated limitations

Additional limitations evident in the study