Ventricular PFA

Definition

Ventricular pulsed field ablation is the delivery of electroporation energy to ventricular myocardium to treat ventricular tachycardia, premature ventricular complexes or ventricular fibrillation. It differs from atrial PFA in one decisive respect: the ventricular wall is thick and often scarred or fatty, and PFA's lesion depth plateaus at roughly 6–8 mm regardless of waveform or catheter design — below what is needed for transmurality across much of the left ventricle. (sources/pfa-af-ehra-2026, rating: very high) Every clinical and technical development in this field is an attempt to escape that ceiling: larger tips, higher voltage, repeated applications, epicardial delivery, sequential RF-plus-PF, and bipolar ablation between two catheters.

Key Concepts

The mechanism, tissue selectivity, device classification and platform specifications are common to atrial and ventricular use and are set out in concepts/PFA-Atrial-Arrhythmias. What follows is specific to ventricular delivery.

Why the ventricle is different

First clinical experience — off-label atrial catheters

Three cases treated with the 31-mm pentaspline Farawave catheter at 2.0 kV bipolar dosing:

Electrogram amplitude fell 30% after one application and 56% after a full set — a modest attenuation consistent with a non-transmural lesion. Repetition dependence was evident in all three cases, and VT termination during PFA may reflect capture of a large myocardial mass rather than circuit elimination. (All: sources/pfa-vt-case-series-jacep-2023)

Preclinical strategies to increase depth

In 14 swine using the 9-mm lattice tip with repeated 5.5-second applications:

Target Technique Dimensions (L × W × D) Transmurality
Papillary muscle (stable contact, n=9) Monopolar, 3–5 applications 18.3 × 15.3 × 5.8 mm Confined to muscle; chordae intact, no MR at 21 days
Papillary muscle (intermittent contact, n=4) Monopolar 18.8 × 12.7 × 3.9 mm (P=0.014 vs stable) Extension into adjacent myocardium
Moderator band (n=2) Monopolar, 3 and 5 applications 19.5–24.0 × 13.5–15.3 × 5.1–6.1 mm Transmural, including insertion sites
Epicardium (n=10) Monopolar, 4–5 applications 30.4 × 23.5 × 9.1 mm 2/10 (20%)
Septum / LV free wall (n=12) Bipolar, two tips, 4–7 applications 29.6 × 21.0 × 14.3 mm 13/15 (87%; P=0.002 vs epicardial)

Dual-energy lattice-tip catheter — first human series

The RF mode of this tip was characterised first: against a conventional 3.5-mm irrigated catheter it doubled ventricular lesion depth in sheep and swine and reached 80% transmurality in bipolar configuration, and it was on that lesion-depth rationale that the catheter entered clinical use. See entities/Sphere-9. (sources/lattice-tip-rf-lesion-circep-2019, rating: medium) (sources/lattice-tip-rf-lesion-jce-2019, rating: medium)

In 18 patients (94% structural heart disease, LVEF 34%, 66% prior failed ablation, 22% electrical storm), high-energy RF was the primary energy and PF consolidated lesions, never near the conduction system and as the only energy epicardially:

The AVAAR registry — largest systematic dataset

126 patients at all 18 European centres using the AFFERA system (VT 99, PVC 23, VF 4; ischaemic 40%, non-ischaemic 46%; LVEF 37%; 64% redo; epicardial access 17%; PFA only 43%, RF only 15%, both 42%):

High-voltage focal PFA

A dedicated focal high-voltage catheter (5 QRS-gated applications of <200 ms, monophasic >10 kV) produced 12.5 ± 3.3 mm deep lesions in swine and, in the first-in-human VCAS trial of 26 patients, 92% acute success, 93% non-inducibility and 81.8% 180-day freedom from VT/VF or ICD shock, with endocardial-only delivery abolishing adjacent epicardial potentials in 9/9. This is single-arm, 6-month evidence only. See concepts/VT-Ablation and sources/pfa-vt-circ-2025. (sources/pfa-af-ehra-2026, rating: very high)

Other ventricular and outflow data summarised in the 2026 statement

Papillary muscle and moderator band targets

The preclinical data are the most encouraging part of the field: papillary lesions of ~6 mm with intact chordae and no mitral regurgitation at 21 days, and transmural moderator band ablation including insertion sites — targets where stable contact has always been the obstacle with thermal catheters. Human evidence remains confined to case reports and the PVC subgroup of registries. (sources/lattice-tip-preclinical-circep-2024, rating: medium)

Contradictions / Open Questions

Connections

Sources