Optimal Pace Timing for LBBAP With or Without an Additional LV Lead: Results From the CSPOT Study

Authors, Journal, Affiliations, Type, DOI

Overview

This CSPOT analysis examined how sensitive LBBAP, LOT-CRT and BVP are to atrioventricular (AV) and interventricular (VV) timing in 48 CRT-indicated patients with preserved AV conduction. All three modalities were similarly sensitive to AV delay: a 60 ms deviation from the optimum reduced the haemodynamic benefit by about 30%. The optimal paced AV delay differed by modality when judged by LV dP/dtmax — LBBAP 152 ms, LOT-CRT 171 ms, BVP 167 ms — but was similar when judged by QRS duration (164, 160 and 165 ms). Optimal AV delay was linearly related to the patient's intrinsic AV conduction: PAV ≈ 0.54 × Ap–LBBs + 3.8 ms for LBBAP, 0.56 × Ap–LBBs + 7.7 ms for LOT-CRT, and 0.48 × Ap–RVs + 28.0 ms for BVP. Patient-specific AV delays derived from these equations outperformed both nominal and best-fixed delays. Simultaneous biventricular activation (VV = 0) was optimal for LOT-CRT and BVP.

Keywords

Cardiac resynchronization therapy; Left bundle branch area pacing; Left bundle branch area pacing-optimized cardiac resynchronization therapy; AV delay; VV delay

Key Takeaways

Design and population

Sensitivity to AV delay

Optimal AV delay

Relationship to intrinsic AV conduction

Fixed versus patient-specific AV delay

Optimal VV delay

Interpretation

Limitations of the document

Key Concepts Mentioned

Key Entities Mentioned

Wiki Pages Updated