Sphere-9 (AFFERA Lattice-Tip Ablation System)

Details

Sphere-9 is the ablation catheter at the centre of the AFFERA electroanatomical mapping and ablation system, developed by Affera (Watertown, MA) and now owned by Medtronic. It is an 8-Fr bidirectional deflectable catheter whose distal electrode is an expandable nitinol lattice that opens into a 9-mm sphere. The lattice carries nine surface sensors that serve both as microelectrodes and as thermocouples, arranged in two rings of four plus one at the tip about 5 mm apart, with a central protected irrigation nozzle. Because blood and saline conduct current away from the metal, the lattice presents an effective ablating surface area of ≈275 mm² — roughly ten times that of a 3.5-mm irrigated tip — and a central electrode that by design never contacts tissue, which helps separate near-field from far-field electrograms. (sources/lattice-tip-rf-lesion-circep-2019, rating: medium)

The system is dual-energy: a unified generator and control architecture switches between high-current temperature-controlled radiofrequency and high-voltage pulsed field trains, with impedance, temperature and voltage feedback governing delivery. The catheter maps with the same tip it ablates with, so lesions can be created at the exact site of signal acquisition without a catheter exchange. Approvals cover the EU and USA. (sources/pfa-af-ehra-2026, rating: very high)

Under the 2026 consensus classification the device is a large-tip focal catheter (largest tissue-contact dimension >4 and ≤12 mm). (sources/pfa-af-ehra-2026)

Key Facts

The RF mode — why the geometry matters

RF lesion performance versus a conventional irrigated tip

Model Setting Lattice Conventional Comparator
Ex vivo bovine/porcine, unipolar, 30 s Tmax 60 °C vs 40 W 6.7 ± 1.3 mm deep 4.8 ± 1.2 mm CF 10 g
Ex vivo, unipolar, 60 s " 8.3 ± 1.4 mm 5.4 ± 0.8 mm
Ex vivo, unipolar, 120 s " 10.0 ± 1.6 mm, 2.4 ± 1 mL 6.1 ± 1.6 mm, 0.5 ± 0.3 mL
Ex vivo, bipolar, 60 s two lattice tips 15.8 ± 4.1 mm, 80% transmural 10.5 ± 1.4 mm, 0% transmural
In vivo swine ventricle, 60 s " 10.5 ± 1.4 mm deep, 27.3 ± 5.2 mm wide 6.5 ± 0.8 mm, 9.2 ± 1.3 mm
Ovine LV, unipolar, 60–120 s 2.7 A / 60 °C / 30 mL/min 10.3 ± 2.9 mm deep, 1,707 ± 771 mm³ 5.5 ± 2.4 mm, 563 ± 398 mm³ 30–35 W / 60 s

(All: sources/lattice-tip-rf-lesion-circep-2019, rating: medium) (sources/lattice-tip-rf-lesion-jce-2019, rating: medium)

The PF mode

Clinical performance in ventricular arrhythmia

Study Design Key results
Peichl 2024 Prospective, 18 patients, 66% prior failed ablation, 22% electrical storm ~4,500 mapping points in ~18 min; 89% acute non-inducibility; 78% ventricular-arrhythmia-free at 3 months; coronary spasm in 3 of 4 epicardial patients; no phrenic palsy; hsTnT 36 → 995 ng/L
AVAAR 2025 Multicentre registry, 126 patients at 18 centres, 64% redo 84% acute success; 72% free of recurrence at 5.6 months (de novo 80% vs redo 68%); 6% major acute complications and 6% major events day 1–30 including 3 deaths

(sources/lattice-tip-vt-europace-2024, rating: medium) (sources/avaar-registry-europace-2025, rating: high)

Atrial use

Contradictions / Open Questions

Connections

Sources